A signal conditioning module for denoising Electrocardiogram signals

dc.contributor.authorPatel, Vandanaeng
dc.contributor.authorShah, Ankiteng
dc.date.accessioned2023-06-28 00:00:00
dc.date.accessioned2025-05-21T19:15:46Z
dc.date.available2023-06-28 00:00:00
dc.date.issued2023-06-28
dc.description.abstractIn this work, we propose to use an optimal multiband filter with least mean square algorithm to design a signal conditioning module for denoising Electrocardiogram (ECG) signals contaminated with predominant noises. The module is implemented on a Field Programmable Gate Array (FPGA) hardware. The experimental results of the proposed module are investigated and compared using an ECGID database available on Physionet. Quantitative and qualitative analysis is performed using Signal to Noise Ratio (SNR), Mean Square Error (MSE), and quality indexes to assess the effectiveness of the module. The average values of SNR are 10.90124, and MSE is 0.001761, indicating the successful elimination of noises in the filtered ECG signal using the proposed module. The signal quality indexes also demonstrate that the relevant information for diagnosing cardiac functionality is preserved. Furthermore, the performance of the designed module is tested on ECG signals obtained from electrodes placed on the human body. The Spartan 3s500efg320-5 FPGA device is employed to implement the filter design module using the partial serial architecture.eng
dc.format.mimetypeapplication/pdfeng
dc.identifier.doi10.32397/tesea.vol4.n1.506
dc.identifier.eissn2745-0120
dc.identifier.urihttps://hdl.handle.net/20.500.12585/13509
dc.identifier.urlhttps://doi.org/10.32397/tesea.vol4.n1.506
dc.language.isoengeng
dc.publisherUniversidad Tecnológica de Bolívareng
dc.relation.bitstreamhttps://revistas.utb.edu.co/tesea/article/download/506/377
dc.relation.citationeditionNúm. 1 , Año 2023 : Transactions on Energy Systems and Engineering Applicationseng
dc.relation.citationendpage67
dc.relation.citationissue1eng
dc.relation.citationstartpage56
dc.relation.citationvolume4eng
dc.relation.ispartofjournalTransactions on Energy Systems and Engineering Applicationseng
dc.relation.referencesSyed Anas Imtiaz, James Mardell, Siavash Saremi-Yarahmadi, and Esther Rodriguez-Villegas. Ecg artefact identification and removal in mhealth systems for continuous patient monitoring. Healthcare technology letters, 3(3):171–176, 2016. [2] Mohammed Mujahid Ulla Faiz and Izzet Kale. Removal of multiple artifacts from ecg signal using cascaded multistage adaptive noise cancellers. Array, 14:100133, 2022. [3] Navdeep Prashar, Meenakshi Sood, and Shruti Jain. Design and performance analysis of cascade digital filter for ecg signal processing. International Journal of Innovative Technology and Exploring Engineering, 8(8):2659–2665, 2019. [4] Kiran Kumar Patro and P Rajesh Kumar. De-noising of ecg raw signal by cascaded window based digital filters configuration. 2015 IEEE Power, Communication and Information Technology Conference (PCITC), pages 120–124, 2015. [5] Inderbir Kaur, Rajni Rajni, and Anupma Marwaha. Ecg signal analysis and arrhythmia detection using wavelet transform. Journal of The Institution of Engineers (India): Series B, 97(4):499–507, 2016. [6] Manas Rakshit and Susmita Das. An efficient ecg denoising methodology using empirical mode decomposition and adaptive switching mean filter. Biomedical signal processing and control, 40:140–148, 2018. [7] Shailesh Kumar, Damodar Panigrahy, and PK Sahu. Denoising of electrocardiogram (ecg) signal by using empirical mode decomposition (emd) with non-local mean (nlm) technique. biocybernetics and biomedical engineering, 38(2):297–312, 2018. [8] Mohamed Sraitih and Younes Jabrane. A denoising performance comparison based on ecg signal decomposition and local means filtering. Biomedical Signal Processing and Control, 69:102903, 2021. [9] Ge Wang, Lin Yang, Ming Liu, Xin Yuan, Peng Xiong, Feng Lin, and Xiuling Liu. Ecg signal denoising based on deep factor analysis. Biomedical Signal Processing and Control, 57:101824, 2020. [10] N Sasirekha, P Vivek Karthick, T Premakumari, J Harirajkumar, and S Aishwarya. Noise removal in ecg signal using digital filters. European Journal of Molecular & Clinical Medicine, 7(2):5145–5149, 2020. [11] C Venkatesan, P Karthigaikumar, and R Varatharajan. Fpga implementation of modified error normalized lms adaptive filter for ecg noise removal. Cluster Computing, 22:12233–12241, 2019. [12] Aya N Elbedwehy, Mohy Eldin Ahmed Abo-Elsoud, and Ahmed Elnakib. Ecg denoising using a single-node dynamic reservoir computing architecture. MEJ. Mansoura Engineering Journal, 46(4):47–52, 2021. [13] Shubhankar Saxena, Rohan Jais, and Malaya Kumar Hota. Removal of powerline interference from ecg signal using fir, iir, dwt and nlms adaptive filter. 2019 International conference on communication and signal processing (ICCSP), pages 0012–0016, 2019. [14] HU Nonyelu, CC Okezie, and HN Uzo. Reducing powerline interference in electrocardiogram signal using optimized trapezoid window based digital finite impulse response filter. World Scientific News, 166:116–131, 2022. [15] Md-Billal Hossain, Syed Khairul Bashar, Jesus Lazaro, Natasa Reljin, Yeonsik Noh, and Ki H Chon. A robust ecg denoising technique using variable frequency complex demodulation. Computer methods and programs in biomedicine, 200:105856, 2021. [16] Kirti Tripathi, Harsh Sohal, and Shruti Jain. Design and implementation of robust low power ecg pre-processing module. IETE Journal of Research, pages 1–7, 2020. [17] Nooraisyah N Samsudin, Suhaila Isaak, and Norlina Paraman. Implementation of optimized low pass filter for ecg filtering using verilog. Journal of Physics: Conference Series, 2312(1), 2022. [18] Harsh Sohal, Shruti Jain, et al. Fpga implementation of collateral and sequence pre-processing modules for low power ecg denoising module. Informatics in Medicine Unlocked, 28:100838, 2022. [19] Amit Singhal, Pushpendra Singh, Binish Fatimah, and Ram Bilas Pachori. An efficient removal of power-line interference and baseline wander from ecg signals by employing fourier decomposition technique. Biomedical Signal Processing and Control, 57:101741, 2020. [20] Goldberger AL, Amaral LAN, Glass L, Hausdorff JM, Ivanov PCh, Mark RG, Mietus JE, Moody GB, Peng C-K, and Stanley HE. Physionet: Components of a new research resource for complex physiologic signals. PhysioBank, PhysioToolkit, and PhysioNet, 2003. [21] Vandana Patel and Ankit Shah. Digital multiband filter design with power spectrum analysis for electrocardiogram signals. 2021 6th International Conference on Recent Trends on Electronics, Information, Communication and Technology, pages 923–927, 2021. [22] John G Proakis and Dimitris G Manolakis. Digital Signal Processing. Prentice Hall, 1995. [23] C Sidney Burrus. Multiband least squares fir filter design. IEEE transactions on signal processing, 43(2):412–421, 1995. [24] James J Bailey, Alan S Berson, Arthur Garson Jr, Leo G Horan, Peter W Macfarlane, David W Mortara, and Christoph Zywietz. Recommendations for standardization and specifications in automated electrocardiography: Bandwidth and digital signal processing. a report for health professionals by an ad hoc writing group of the committee on electrocardiography and cardiac electrophysics. Circulation, 81(2):730–739, 1990. [25] Ivan Dotsinsky. Review of advanced methods and tools for ecg data analysis. BioMedical Engineering OnLine, 6(1):1–18, 2007. [26] Thion Ming Chieng, Yuan Wen Hau, Zaid Bin Omar, and Chiao Wen Lim. Qualitative and quantitative performance comparison of ecg noise reduction and signal enhancement method based on various digital filter designs and discrete wavelet transform. International Journal of Computing and Digital Systems, 9(4):553–565, 2020.eng
dc.rightsVandana Patel, Ankit Shah - 2023eng
dc.rights.accessrightsinfo:eu-repo/semantics/openAccesseng
dc.rights.coarhttp://purl.org/coar/access_right/c_abf2eng
dc.rights.creativecommonsThis work is licensed under a Creative Commons Attribution 4.0 International License.eng
dc.rights.urihttps://creativecommons.org/licenses/by/4.0eng
dc.sourcehttps://revistas.utb.edu.co/tesea/article/view/506eng
dc.subjectSignal Conditioningeng
dc.subjectElectrocardiogrameng
dc.subjectField Programmable Gate Arrayeng
dc.subjectMultiband filtereng
dc.titleA signal conditioning module for denoising Electrocardiogram signalsspa
dc.title.translatedA signal conditioning module for denoising Electrocardiogram signalsspa
dc.typeArtículo de revistaspa
dc.type.coarhttp://purl.org/coar/resource_type/c_6501eng
dc.type.coarversionhttp://purl.org/coar/version/c_970fb48d4fbd8a85eng
dc.type.contentTexteng
dc.type.driverinfo:eu-repo/semantics/articleeng
dc.type.localJournal articleeng
dc.type.versioninfo:eu-repo/semantics/publishedVersioneng

Archivos