Estimation of PQ distance dispersion for atrial fibrillation detection
Fecha
2021-05-03Autor(es)
Giraldo-Guzmán J.
Kotas, Marian
Castells, Francisco
Contreras Ortiz, Sonia Helena
Urina-Triana, Miguel
Metadatos
Mostrar el registro completo del ítemResumen
Background and objective: Atrial fibrillation (AF) is the most common cardiac arrhythmia in the world. It is
associated with significantly increased morbidity and mortality. Diagnosis of the disease can be based on
the analysis of the electrical atrial activity, on quantification of the heart rate irregularity or on a mixture
of the both approaches. Since the amplitude of the atrial waves is small, their analysis can lead to false
results. On the other hand, the heart rate based analysis usually leads to many unnecessary warnings.
Therefore, our goal is to develop a new method for effective AF detection based on the analysis of the
electrical atrial waves. Methods: The proposed method employs the fact that there is a lack of repeatable P waves preceding QRS complexes during AF. We apply the operation of spatio-temporal filtering (STF) to magnify and detect the prominent spatio-temporal patterns (STP) within the P waves in multi-channel ECG recordings. Later we
measure their distances (PQ) to the succeeding QRS complexes, and we estimate dispersion of the obtained PQ series. For signals with normal sinus rhythm, this dispersion is usually very low, and contrary, for AF it is much raised. This allows for effective discrimination of this cardiologic disorder. Results: Tested on an ECG database consisting of AF cases, normal rhythm cases and cases with normal rhythm restored by the use of cardioversion, the method proposed allowed for AF detection with the accuracy of 98.75% on the basis of both 8–channel and 2–channel signals of 12 s length. When the signals length was decreased to 6 s, the accuracy varied in the range of 95% − 97.5% depending on the number of channels and the dispersion measure applied. Conclusions: Our approach allows for high accuracy of atrial fibrillation detection using the analysis of electrical atrial activity. The method can be applied to an early detection of the desease and can advantageously be used to decrease the number of false warnings in systems based on the analysis of the heart rate.
Citar como
Jader Giraldo-Guzmán, Marian Kotas, Francisco Castells, Sonia H. Contreras-Ortiz, Miguel Urina-Triana,Estimation of PQ distance dispersion for atrial fibrillation detection, Computer Methods and Programs in Biomedicine, Volume 208, 2021, 106167, ISSN 0169-2607, https://doi.org/10.1016/j.cmpb.2021.106167.Utilice esta dirección para citar:
https://hdl.handle.net/20.500.12585/10368Colecciones
- Productos de investigación [1453]
Compatible para recolección con:
Archivos
El ítem tiene asociados los siguientes ficheros de licencia:
Universidad Tecnológica de Bolívar - 2017 Institución de Educación Superior sujeta a inspección y vigilancia por el Ministerio de Educación Nacional. Resolución No 961 del 26 de octubre de 1970 a través de la cual la Gobernación de Bolívar otorga la Personería Jurídica a la Universidad Tecnológica de Bolívar.