Development of an online optical prototype for the simultaneous analysis of defects or gels in industrially synthesized polypropylene films. Part 1: Comparison with ASTM D 3351–93 methods
Fecha
2024-04-01Autor(es)
Hernandez Fernandez, Joaquin
Ortiz, Katherine
Lopez-Martinez, Juan
Metadatos
Mostrar el registro completo del ítemResumen
Polypropylene (PP) films are crucial in various industrial applications, from packaging to medical prod ucts. However, a common challenge in PP manufacturing is the presence of gel-like defects. These gels
are minor defects on the surface of the films that significantly affect the physicochemical, mechani cal, and organoleptic properties of the films, compromising the quality of the final product. This first
research focuses on developing and validating an in-line optical method to replace the international
method ASTM D 3351–93. The main objective was to create a methodology that has the same scope
and analytical performance as those reported by ASTM D 3351–93 in such a way that it can compete
with it in terms of precision and accuracy, thus allowing end users to this ASTM, such as PP producers,
PP marketers, PP film producers, among others internationally, can use this new methodology with
necessary analytical support. This analytical methodology integrates the PP extrusion zones, the film
processing stages, and the optical zone for reading and processing analytical data. Additionally, it has
the advantage of working with a sample size that is even more representative of the population and
has less human error since only one operator is required to carry out the test; this method also has
much shorter response times. The developed prototype had 14 online stages that allowed represen tative quantities of samples to be taken and processed thermally and mechanically for ideal optical
measurement. For the online method, a 6-point calibration curve is carried out at concentrations of 40,
10, 5, 2, 1 and 0 ppm for the gel or defect sizes of 200, 400, 500, 600, 700, 800 and 900 μm, show ing excellent linearity where the correlation coefficient varied between 0.997 and 0.999, the limits of
detection (LOD) varied between 0.85 and 2.61 and the limits of quantification (LOQ) ranged between
2.82 and 8.71. The statistical analyzes by ANOVA of the comparison between the ASTM D 3351–93
method and the proposed simultaneous method indicate that the p value of the evaluation of the means
was 0.946, which suggests that the means are not statistically different. To complement, the Tukey test
was carried out at 95 %, indicating that the methods have statistical equivalence.
Citar como
Joaquina, H.-F., Katherine, O., & Juan, L.-M. (2024). Development of an online optical prototype for the simultaneous analysis of defects or gels in industrially synthesized polypropylene films. Part 1: Comparison with ASTM D 3351–93 methods. MethodsX.Utilice esta dirección para citar:
https://hdl.handle.net/20.500.12585/12671Colecciones
- Productos de investigación [1453]
Compatible para recolección con:
Archivos
El ítem tiene asociados los siguientes ficheros de licencia:
Universidad Tecnológica de Bolívar - 2017 Institución de Educación Superior sujeta a inspección y vigilancia por el Ministerio de Educación Nacional. Resolución No 961 del 26 de octubre de 1970 a través de la cual la Gobernación de Bolívar otorga la Personería Jurídica a la Universidad Tecnológica de Bolívar.